Product Details
HC-9884 electrolyte analyzer uses international advanced ion-selective electrode. The detection sensitivity of the machine for Potassium, Sodium, Chloride electrolyte, stability, anti-interference ability reached the advanced level of similar foreign models. The Chinese operational information management system of the machine meets the needs of the medical unit. The equipment uses the technology of pressure sensor ingenious in detection of CO2. Both to ensure the accuracy of the test, and greatly simplifies the complex steps in the reference method. The equipment uses advanced microprocessor control system to achieve a fully automated test. Potassium, sodium, chloride, total CO2, AG(anion gap ) and other projects can be detected simultaneously.
Test range:
Analyte
Units
Range
Resolution
K+
mmol/L
0.40
15.00
0.01
Na+
mmol/L
30.0
200.0
0.1
Cl-
mmol/L
30.0
200.0
0.1
TCO2
mmol/L
4.0
100.0
0.1
AG
mmol/L
Calculate
0.1
Technical Specification:
Sample Types:
Blood , Serum , Plasma or Diluted Urine
Sample Size:
60ul ~ 220ul serum , urine
Analysis Time:
25 ~ 45 seconds serum (including 1-point calibration) (HC9883 & HC9885)
35 ~ 60 seconds serum (including 1-point calibration) (HC9884 & HC9886)
Throughput:
80 ~ 110 samples per hour
Power Source:
AC220V, 50Hz +/- 10V
Power Consume:
< 60W
Instrument Size:
420mm(H) x 240mm(W) x 360mm(L)
Package Size:
630mm(L) x 305mm(H) x 460mm(W)(FORMER)
695mm(L) x 320mm(H) x 480mm(W)(NEW)Environmental Temperature
Range:10 ~ 35℃
Measurement Technology:
Direct Potentiometry
Reagent Include:
Calibrant A, Calibrant B , Electrode Cleanning Solution , QC Solution ,
Internal Filling Solution for Electrode etc.